How long does the RO have to be running before we perform any water testing?
15 minutes
What does IFU stand for?
instructions for
... [Show More] use
***Instructions written by the manufacturer that are approved by the FDA.
How long does the sample valve need to be flushed after the RO has been running for 15 minutes before a sample is collected?
2 minutes
What is the frequency for testing the Total Chlorine?
Before the first patient of the day is put on AND every 4 hours after that.
Name the test trip used to test Total Chlorine:
RPC Ultra-Low Total Chlorine Strips
What does TDS stand for?
Total Dissolved Solids
The RPC Ultra-Low Total Chlorine Strips are used to test the functionality of what tanks in the water room?
GAC (Granular Activated Carbon) tanks
Which tanks in the water room are responsible for removing Total Chlorine?
GAC tanks (Granular Activated Carbon tanks)
What is the TDS meter used to measure?
The TDS meter measures the functionality of the RO and how well it is removing those particles we do not want to expose the patient to during treatment.
What 3 things does the pHoenix Meter, pHoenix XL Meter and Myron L-D6 Meter test dialysate for?
1. conductivity
2. pH
3. temperature
What strips are used to measure water hardness?
SteriChek Water Hardness Reagent Strips
What do the SteriChek Water Hardness Reagent Strips measure for?
water hardness
How long does the RO have to be running before we perform any water testing?
15 minutes
What does IFU stand for?
instructions for use
***Instructions written by the manufacturer that are approved by the FDA.
How long does the sample valve need to be flushed after the RO has been running for 15 minutes before a sample is collected?
2 minutes
What is the frequency for testing the Total Chlorine?
Before the first patient of the day is put on AND every 4 hours after that.
Name the test trip used to test Total Chlorine:
RPC Ultra-Low Total Chlorine Strips
What does TDS stand for?
Total Dissolved Solids
The RPC Ultra-Low Total Chlorine Strips are used to test the functionality of what tanks in the water room?
GAC (Granular Activated Carbon) tanks
Which tanks in the water room are responsible for removing Total Chlorine?
GAC tanks (Granular Activated Carbon tanks)
What is the TDS meter used to measure?
The TDS meter measures the functionality of the RO and how well it is removing those particles we do not want to expose the patient to during treatment.
What 3 things does the pHoenix Meter, pHoenix XL Meter and Myron L-D6 Meter test dialysate for?
1. conductivity
2. pH
3. temperature
What strips are used to measure water hardness?
SteriChek Water Hardness Reagent Strips
What do the SteriChek Water Hardness Reagent Strips measure for?
water hardness [Show Less]